Calibration of different Instruments
Calibration of pH meter:
- · Calibrate the pH meter with buffer solution
- · Check the pH electrode
- · Examine the electrode for contamination or damage
- · Flush the pH sensor with distilled water
- · Dip the electrode in buffer solutions
- · Get different pH values such as 3, 7 and 9
- · Wash the pH sensor after every determination
- · Now our pH meter is calibrated
Calibration of HPLC:
Calibration of HPLC depend on different parameters:
Flow rate accuracy:
- · Prime all the solvents with water
- · Set the flow rate 0.500 ml/m
- · Wait for 15 minute to stabilize the system
- · Stabilize the pressure
- · Start the stop watch after inserting outlet tube into 10ml volumetric flask
- · When the lower meniscus reaches 10ml on flask, stop the watch
- · Record the elapsed time and check flow rate
System precision:
- · Inject the blank
- · Note the areas and retention time
- · Calculate %RSD of retention time and peak areas
Wavelength accuracy:
- · Create an instrument method with a wavelength in nm
- · Inject the blank
- · Note down the height and absorbance
Detector linearity:
- · Inject the blank followed by detector linearity solutions
- · Record the peak responses plot between concentration and peak response
Injector linearity:
- · Inject 5 µl of mobile phase as an blank injection
- · Inject different mobile phases
- · Record the areas
- · Plot a curve for volume injected Vs peak area
Gradient performance
check:
- · Add 5ml acetone to 1000ml of methanol.
- · Connect the pump and detector inlet with union
- · Set the detector wavelength at 254nm
- · Set the gradients programme with flow rate such as A, B, C and D
Column oven
temperature accuracy:
- · Connect the column to detector
- · Place the thermometer probe in column oven and set the column oven temperature at 40degree and wait to stabilize it
- · Note the temperature
- · Repeat it on different temperatures
Calibration of spectrophotometer:
- · Warm
up the spectrophotometer for 45 minutes.
- · Select
the wavelength to calibrate.
- · Depending
on what type of filter you are, using will determine if you need a dedicated
blank for the machine. Either way, if you are standard came with a blank
insert it at this time, if there is no blank leave the cuvette holder empty and
close the lid.
- · Zero
the spectrophotometer so the display shows all zeros.
- · Remove
the blank (if you used one).
- · Insert
the NIST calibration standard and close the cover.
- · Record
the reading and compare it to the data on the certificate of calibration.
- · If
data will match then it will be done
Calibration
of Viscometer:
- · Ensure that instrument is ready for use.
- ·
Calibrate
the instrument by using different calibration standards solutions in 500ml
beaker
- ·
As
per requirement, we need to select the spindle by pressing “S" key on
display.
- ·
Pressing
“up or down" arrow keys on display while blinking
- ·
Two
type digits will change the value accordingly, “increase or decrease no. of
spindle"
- ·
We
also need to select the speed of spindle as per requirement by pressing “Set
Speed" key
- ·
RPM
speed also select as per requirement by pressing “up or down” keys
- ·
Safely
attach the selected spindle to the lower shaft
- ·
Lift
it slightly and holding it firmly with one hand while screwing the spindle on
the other hand
- ·
Put
the test solution beaker below the spindle center and insert the spindle until
the fluid level is at immersion groove in the spindle shaft middle of the
“Spindle mark line"
- ·
Press
the motor key to “on or off" to run the test
- ·
Allow
the time to stabiles the reading.
- ·
Record
the stabilized reading in CPS unit in calibration log
- ·
Press
the key, “on or off" to stopping the test and putting “off" the
instrument mains.
- · Clean the spindle and instrument properly before and after use
Calibration of Hydrometer:
- · We need to understand the basic principle of hydrometer is an instrument used to determine specific gravity
- · It operates based on the Archimedes principle that a solid body displaces its own weight within a liquid in which it floats.
- · Each
hydrometer is cleaned by a thorough rinse with ethanol
- · It
is dried with a clean, paper towel before it is calibrated
- · Hydrometer
calibration by Cuckow’s method requires that the hydrometer be weighed in air
- · And weighed while immersed to the selected scale marks in a liquid of known density
Calibration of Texture Analyzer
- When we need to calibrate the analyzer must ensure that our texture analyzer should be on which button located in the left back corner
- Turn on the computer
- Before start testing, you need to first calibrate the instrument
- To do this, attach the probe as you like and start calibration
- From the calibration menu, select the calibration force
- You need to use the calibration weight and then enter the calibration weight value into grams
- When it's done, remove the weight
- Next to calibration the height, you need to select the probe as you like
- Run it till the probe touches platform and then back up
- Texture analyzer is ready now for test
Calibration
of Gas chromatography (GC):
- Basically
- · Some basic steps of gas chromatography first we need to switch the GC instrument
- · And then open the gas cylinder after that
- · Adjust pressure and then check the carrier gas flow rate
- · The GC calibration ensure that the information of measured signal directly correspond to the bulk and specific properties
Calibration
of Refractometer:
- · First, we need a calibration liquid we also use distilled water
- · Place the plate in daylight then add 2 3 drops of calibration liquid
- · Then we need to close the light and allow to calibrate liquid and spread around the prism with no dry spot then wait for a little bit
- · After the wait sample reaches the temperature of the refractometer
- · Calibrate the device using distilled water and ensure the refractometer surface should be dry and clean
Calibration of Flame photometer:
- · We
need to understand the basic principle of flame photometer like concentration
ranges and light
- · In
flame photometer we just determine the concentration of certain solutions
- · The
most important part of calibrating the flame photometer is we have a point of
reference for a reading then control the entire testing process include flame
photometer, sample and operator
No comments:
Post a Comment